Line 39: |
Line 39: |
| [[File:Clifford.png|thumb|Instrumentation Technologist Clifford Chouinor sets up working standard for calibration at the CNSC laboratory]] | | [[File:Clifford.png|thumb|Instrumentation Technologist Clifford Chouinor sets up working standard for calibration at the CNSC laboratory]] |
| | | |
− | ==== What were the challenges the laboratory faced while preparing for accreditation? ==== | + | ==== What were the challenges the laboratory faced while preparing for accreditation? ==== |
| + | Adapting all in-house MS Excel spreadsheets, writing procedures and work instructions to comply with all the ISO/IEC 17025:2005 standard requirements was certainly a big challenge. There are a lot of documents! |
| + | |
| + | Another significant challenge was to find proficiency testing (PT) providers with appropriate measurement capabilities to meet ISO compliance requirements. The capabilities of the few established and reputable PT providers, such as the National Research Council of Canada (NRC), the National Institute of Standards and Technologies (NIST) and the International Atomic Energy Agency (IAEA), are limited to their well-characterized artefacts consisting of working measurement instruments for air kerma measurements. Their PT capabilities do not extend to field instruments such as personal dosimeters and gamma survey equipment. |
| + | |
| + | It is important to note that the accreditation task presented to the Standards Council of Canada (SCC) and NRC CLAS, with the measurement scope that the CNSC laboratory sought, was the first of its kind for all three of these primary project stakeholders, and a first for Canada. The unique scope of accreditation turned out to be a challenge for the SCC and the NRC Calibration Laboratory Assessment Service (NRC CLAS), as the requirements for the analysis of measurement uncertainties and model equations, in addition to data analysis of the measurements and interpretation of the comparison of results from PT exercises are different from other physical and electrical metrological calibrations. |
| | | |
| ===== Typical instruments that the CNSC laboratory calibrates under the scope of accreditation ===== | | ===== Typical instruments that the CNSC laboratory calibrates under the scope of accreditation ===== |
Line 49: |
Line 54: |
| ==== How can CNSC staff members suggest improvements to the laboratory? ==== | | ==== How can CNSC staff members suggest improvements to the laboratory? ==== |
| | | |
− | Through annual customer satisfaction surveys that the CNSC laboratory conducts, CNSC staff members who have used the laboratory services in the previous year can provide feedback on their quality, including suggestions for improvement.The CNSC laboratory conducted such surveys in 2015 and 2016, and has taken action to address comments from its customers to meet its commitment to provide a high service level and meet customer expectations. One such action was to respond to the suggestion to provide links to instrument manuals from the laboratory’s Web page. | + | Through annual customer satisfaction surveys that the CNSC laboratory conducts, CNSC staff members who have used the laboratory services in the previous year can provide feedback on their quality, including suggestions for improvement. The CNSC laboratory conducted such surveys in 2015 and 2016, and has taken action to address comments from its customers to meet its commitment to provide a high service level and meet customer expectations. One such action was to respond to the suggestion to provide links to instrument manuals from the laboratory’s Web page. |
| | | |
| The CNSC laboratory gamma calibration service now operates within a quality management system that complies with the requirements of ISO/IEC 17025:2005. As specified in the quality management system, the laboratory implements mechanisms including, but not limited to, proficiency testing, customer satisfaction surveys, internal reviews, management reviews, and corrective and preventative actions to drive continual improvement, which is the true intention of the ISO/IEC standard. | | The CNSC laboratory gamma calibration service now operates within a quality management system that complies with the requirements of ISO/IEC 17025:2005. As specified in the quality management system, the laboratory implements mechanisms including, but not limited to, proficiency testing, customer satisfaction surveys, internal reviews, management reviews, and corrective and preventative actions to drive continual improvement, which is the true intention of the ISO/IEC standard. |